PN#: AGG1314B
 |
Series 1200 UV-Visible Detector The 1200 Series variable wavelength detector provides improved detection limits with electronic temperature control, programmable wavelength switching, stop-flow wavelength scanning, and extensive diagnostics and error detection.
Available with full range of flow cells.
Click to view datasheet
|
Noise: ± 0.15 x 10^-5 AU, at 230 nm
Drift: 1 x 10^-4AU/hr, at 230 nm
Linearity: >2.5 AU upper limit
Wavelength Range: 190-600 nm
|
|
PN#: AGG1315D
 |
Series 1200 Diode Array Detector The Series 1200 Diode Array Detector includes deuterium and tungsten lamps along with 1024 diodes and 1-nm slit to provide expanded detection range with high spectral resolution. Unit also features electronic temperature control, reference wavelength, and upgradeability to 80 Hz sampling rate.
Available with 9 analytical and preparative flow cells
Click to view datasheet
|
Short-Term Noise: ± 0.8 x 10^-5 AU at 254 nm and at 750 nm
Drift: 0.9 x 10^-3 AU/h at 254 nm
Linearity: >2 AU upper limit
Wavelength Range: 190-950 nm
|
|
PN#: AGG1321A
 |
Series 1200 Fluorescence Detector The 1200 Series fluorescence detector provides automatic optimization of PMT gain setting and improved selectivity and sensitivity by simultaneous multiple wavelength detection. Detector contains built-in diagnostics tools to minimize downtime; and built-in wavelength calibration for GLP compliance. Click to view datasheet
AGG4208A Instant Pilot Control Module for Stand-alone control.
Click to view datasheet
|
Performance: 10 fg anthracene LOD, Ex 250 nm, Em 400 nm
Pulse Frequency: 365 Hz for signal mode, 73 Hz for economy mode
Spectral Acquisition: Excitation or emission spectra, scan speed 28 ms per data point
Flow Cell: Standard: 8 µL volume, 20 bar (2 MPa)
|
|
BACK |